Toolkit / Neuromodulation

ECT stimulus dose calculator (titration, age-based, half-age)

Work out the treatment stimulus for each electrode placement: from the seizure threshold found at titration (the preferred method), or with the age-based and half-age methods. Shows the dose in millicoulombs and as a Thymatron percentage, and converts stimulus parameters into charge and energy.

Patient details for the printout optional · never saved or sent

Left blank, these print as lines to fill in by pen.

1 Treatment dose

Dosing method
Threshold entered in

2 Stimulus parameters to charge optional

Convert the settings shown on the device into charge and energy, e.g. to compare doses between machines.

ChargeI × PW × f × T × 2–
EnergyQ × I × Z–
Thymatron %Q ÷ 5.04–

The same threshold at other multiples

Enter a seizure threshold to compare multiples.

Usual treatment dose

Right unilateral6× threshold or more (brief pulse); 2.5× is less effective
RUL ultrabriefUsually 6× threshold
Bitemporal1.5–2.5× threshold (1.5× has less cognitive risk)
BifrontalUsually 1–1.5× threshold

Age-based settings (Thymatron)

Age 20Age method (unilateral) 20% = 101 mC · Half-age (bitemporal) 10% = 50 mC
Age 30Age method (unilateral) 30% = 151 mC · Half-age (bitemporal) 15% = 76 mC
Age 40Age method (unilateral) 40% = 202 mC · Half-age (bitemporal) 20% = 101 mC
Age 50Age method (unilateral) 50% = 252 mC · Half-age (bitemporal) 25% = 126 mC
Age 60Age method (unilateral) 60% = 302 mC · Half-age (bitemporal) 30% = 151 mC
Age 70Age method (unilateral) 70% = 353 mC · Half-age (bitemporal) 35% = 176 mC
Age 80Age method (unilateral) 80% = 403 mC · Half-age (bitemporal) 40% = 202 mC

Notes

  • Titration (empirical dosing) is the preferred way to set the dose for every placement: at the first session, stimulate from a low dose and step up, 20–30 seconds apart, until an adequate generalised seizure occurs. That charge is the seizure threshold. Usually no more than 3–4 stimulations are given under one anaesthetic. Use your unit’s titration schedule.
  • Treatment dose as a multiple of the threshold (NVvP 2021, based on Sackeim 2000 and later trials): right unilateral, 6× threshold or more (2.5× is less effective); bitemporal, 1.5–2.5× (starting at 1.5× carries less cognitive risk); bifrontal, usually 1–1.5×. Ultrabrief right unilateral ECT is also usually given at 6× threshold, but may need more sessions.
  • Age-based method (right or left unilateral): Thymatron setting in % = age in years, so dose = age × 5.04 mC. Half-age method (bitemporal): % = half the age (Petrides and Fink 1996). Useful when sub-convulsive stimulation must be avoided, but less precise than titration because thresholds vary widely between people.
  • Fixed high doses for everyone are discouraged for routine use because of greater cognitive side effects.
  • The seizure threshold often rises during a course. If seizures become short or poor in quality, or the patient is not improving, raise the dose to keep the intended multiple, or re-titrate. In continuation or maintenance ECT, start from the last dose that gave an adequate seizure.
  • Many things raise the threshold and shorten seizures: older age, male sex, bitemporal versus unilateral differences, high anaesthetic doses (especially propofol), benzodiazepines, anticonvulsants (including valproate and lamotrigine), and previous ECT. Many lower it: caffeine, theophylline, hyperventilation, and a longer gap since the last session.
  • Charge (mC) = current (A) × pulse width (ms) × frequency (Hz) × train duration (s), multiplied by 2 when the frequency counts bidirectional pulse pairs (as MECTA and Thymatron do). Energy (J) ≈ charge (C) × current (A) × dynamic impedance (Ω); devices show energy at an assumed impedance (220 Ω for the Thymatron). Charge, not energy, is the preferred measure of dose.
  • In India, the Mental Healthcare Act 2017 (section 95) allows ECT only with muscle relaxant and anaesthesia (modified ECT); for minors it needs guardian consent and Mental Health Review Board permission.

References

  1. Nederlandse Vereniging voor Psychiatrie (NVvP). Richtlijn Elektroconvulsietherapie (ECT), module “Doseringsmethode en gebruikte dosis bij ECT”. 2021. richtlijnendatabase.nl.
  2. Ferrier IN, Waite J, eds. The ECT Handbook. 4th ed. Royal College of Psychiatrists / Cambridge University Press; 2019.
  3. American Psychiatric Association. The Practice of Electroconvulsive Therapy: Recommendations for Treatment, Training, and Privileging. 2nd ed. APA; 2001.
  4. Petrides G, Fink M. The “half-age” stimulation strategy for ECT dosing. Convuls Ther. 1996;12(3):138–146.
  5. Abrams R. Electroconvulsive Therapy. 4th ed. Oxford University Press; 2002 (age-based dosing).
  6. Sackeim HA, Prudic J, Devanand DP, et al. A prospective, randomized, double-blind comparison of bilateral and right unilateral electroconvulsive therapy at different stimulus intensities. Arch Gen Psychiatry. 2000;57(5):425–434.
  7. Sackeim HA, Prudic J, Nobler MS, et al. Effects of pulse width and electrode placement on the efficacy and cognitive effects of electroconvulsive therapy. Brain Stimul. 2008;1(2):71–83.
  8. Kellner CH, Knapp R, Husain MM, et al. Bifrontal, bitemporal and right unilateral electrode placement in ECT: randomised trial. Br J Psychiatry. 2010;196(3):226–234.
  9. Kolshus E, Jelovac A, McLoughlin DM. Bitemporal v. high-dose right unilateral electroconvulsive therapy for depression: a systematic review and meta-analysis of randomized controlled trials. Psychol Med. 2017;47(3):518–530.
  10. Peterchev AV, Rosa MA, Deng ZD, Prudic J, Lisanby SH. Electroconvulsive therapy stimulus parameters: rethinking dosage. J ECT. 2010;26(3):159–174.

An educational aid for clinicians, not a substitute for clinical judgement. Follow your ECT unit’s protocol and the device manufacturer’s instructions; the treating team decides the dose. Nothing you enter on this page is stored or sent.